The Dissolution Solutions NetworkCalibrating dissolution test equipment:General tips and helps |
Currently there are no specified calibration intervals in the USP several studies are under way to investigate need and effectiveness of the parameters of calibration. Calibration is currently done by verifying 9 physical parameters and testing two types of tablets. Each of which is shown to be very sensitive to out of specification parameters, defects, or irregularities.
Several proposed changes in calibration or being investigated, such as
defining time of calibration, increasing mechanical calibration, quantifying
mechanical calibrations, and decreasing chemical calibrations. There are many
differing opinions as to changes in dissolution calibration. I find that the
chemical calibrator tablets provide me with a wealth of information that the
mechanical calibration can not even quantify.
Weaknesses of Basket Method
Drug Information Journal, Vol. 30, pp. 1045?1054, 1996 0092-8615/96
There are many factors that will cause a dissolution analysis to give incorrect or errant results. Most dissolution apparatus will not pass calibration unless all conditions are optimal in the apparatus, the standards, the media, and the calibrator tablets. Therefore, it is possible to narrow down all possible causes of failure to calibrate to only a few conditions.
Improper deaeration is the most common problem in dissolution analysis. Heat the media to 41C. Vacuum filter through a 0.45-um filter into a 4 liter vacuum flask, stirring with a magnetic stirrer. Continue to draw a vacuum and stir for an additional 5 minutes. When done properly, this process will eliminate almost all dissolution problems. Media should be deaerated immediately before use. Gently transfer the medium directly to the vessel. Do not introduce air into the medium. Reaeration occurs within 30 minutes. However, excess air only occurs after 20-24 hours.
Vibration of the system while appearing insignificant can affect such factors as shaft perpendicularity, head coplanarity, and tension on the drive belt, centering and operation condition of the gear plates. To minimize vibration effects, the dissolution equipment should be on a stable bench or table. Other mechanical equipment using fans, pumps or other vibration sources should be removed from the area or isolated in some other way. The Bulldozer effect has been shown to give higher results in prednisone. Trains, air conditioning, pumps can cause vibration,
The calibrator tablets should be stored in the original containers in a dry place. Avoid excess humidity. Use the tablets immediately on removal from the container. Do not put the tablets out while waiting to start the test.
Use the current lot of USP reference standard and follow any drying instructions on the label.
Do not centrifuge sample. A separate clean dry filter should be use for each sample and the standard.
Here are the results of some past variability studies done on excess in dissolution on calibrator tablets.
Variable |
Maximum Allowable* |
Effects of excess |
Method of control |
|---|---|---|---|
Wobble |
1.0 mm |
(4-6%) |
replace shafts |
Vibration |
0.1 mil. displacement @ vessel |
5-10% |
Eliminate source |
Alignment |
1.5Degr. to perpendicular |
2-25% |
Adjust alignment |
Centering |
2mm |
(2-13%) |
Adjust centering |
Agitation Rate |
4% |
Linear |
belts, gears |
Dissolved Gas |
Deaerated |
+/-50% |
deaerate media |
Media pH |
varies (usually +/-0.05pH) |
10% |
check buffers, pH meter |
Evaporation |
none |
Linear |
use evaporation covers |
Temperature |
+/-0.5C |
Linear |
Monitor temperature |
Flow |
no interference |
? |
remove resident probes |
*Not all of these are compendial requirements.
Reference: Hanson's Handbook of Dissolution testing: 2nd edition
Pharmacopeial Forum: Volume 20, Number 6