DSN:The Dissolution Solutions Network. Troubleshooting dissolution test and equipment.
Home > Calibration > Mechanical

Physical Calibration

A mechanical calibration protocol study is currently under way to examine whether mechanical calibration is equal to or better than calibrator tablets. However opinions vary in regards to eliminating or reducing chemical calibrators.

Calibration parameter

Current USP tolerances

Point of measurement

Proposed tolerances

Proposed frequency

Height check

25+-2mm

Paddle/basket bottom

25+-2mm

At time of use if vessels are removed

Basket wobble

+-1mm as runout

Bottom of basket rim

+-1.0 mm total run out

Every 3 months

Rotational speed

+-4%

Not applicable

+-1 rpm

Every 3 months

Vessel-shaft centering

2mm from center line

Center line

2mm from center line

Every 3 months

Vessel temperature

37+-0.5C

Not applicable

37+-0.5C

At time of use

Bath levelness

Level

Base plate

Level

Every 3 months

Shaft wobble

no significant wobble

2cm above top of paddle blade

<0.5 mm total run out

Every 3 months

Paddle/basket dimensions

See USP

See USP

See USP

Upon receipt

Vessel dimensions

See USP

See USP

See USP

Upon receipt

Degassing technique

NCDA2 passes calibration

Not applicable

NCDA2 passes calibration

Once per year

Proposed parameters under investigation

 

 

 

 

Paddle/basket examination

None

Not applicable

No gross defects as determined by visual inspection

At time of use

Shaft verticality

None

Above vessel

Vertical as measured by a bubble level

Every 3 months

Visual inspection of belts; check ball bearings

None

Not applicable

Belts are tight, clean, free of cracks, properly aligned; shaft turns freely

Every 3 months

Vibration

None

Base plate while operating at 100 rpm

<0.2mil

Every 3 months

Literature Cited: Pharmacopeial Forum: Volume 26, Number 4