DSN:The Dissolution Solutions Network. Troubleshooting dissolution test and equipment.
Home > Calibration > Prednisone

Prednisone

CONDITIONS:
242nm
500 ml Deionized Water
30 minutes
Absorptivity is usually 44-45

Prednisone tips

Prednisone: a disintegrating tablet. Highly affected by mechanical parameters and vessels. Recently reformulated to be sensitive to poorly deaerated media.

Prednisone, according to the Pharma study, when tested by paddles at 50 RPMs, is operationally sensitive to several perturbations, including: centering, vibration, shaft wobble, shaft verticality, and degassing.

There have been many concerns about the new lot of Prednisone Lot M. Many have stated that their systems that routinely passed, will no longer easily pass. Recently Tahseen Mirza, formerly of the USP, posted suggestions on line. Here are his suggestions to eliminate any problems.

Recently one lab wrote in because they had a problem with prednisone calibration. Construction outside of the facility caused infrequent vibration. The construction effect gave higher result of prednisone during calibration

Another lab wrote in because they had been degassing with helium. They suddenly had a problem with prednisone calibration. They switched to vacuum filtration for degassing and all passed perfectly!

Literature cited: Dissolution Technologies November 2000.
Tahseen Mirza, Q+A: http://www.dissolutiontech.com/DTresour/1100ART/1100Q%26A.htm



Do you have any Prednisone tips that you want to post online?