You may have a problem with calibrating with the new 10mg Prednisone. The lot was reformulated to make it more sensitive to deaeration technique. This is the number one problem that causes the prednisone calibration to be out.

For example, many labs that use helium sparging have found that this technique does not always work with Prednisone 10mg (NCDA2). This is because helium sparging is not really deaerating. Rather, it is replacing nitrogen and oxygen with helium, which has a faster out gassing rate than the other two, but still needs to out gas. So if you sparge then run you may have helium interference. You must determine your rate of out gassing with this technique and determine the appropriate time to start the test.

If you use the USP recommended method of heating to 41C and then vacuum filter while stirring for 5 minutes, you remove almost all of the gas. The media will start to reaerate over the next 30 minutes. This is the point in which it will start to take in excess gas. It is the excess gas that will cause the problems not dissolved gas. So if you use this technique, and run the samples without delay. You should have no problems.

The other major source of error with disintegrating tablets is vessel cleaning. The floating encapsulated micro particles are affected by the hydrodynamic flow and follow an outward sink inward rise flow pattern. This can be hindered if the glass is dirty or even has microfractures or scratches in the glass. Determine if the failure occurs in certain glass vessels by keeping the same vessels in the same location and noting their position.